Generic placeholder image

Combinatorial Chemistry & High Throughput Screening

Editor-in-Chief

ISSN (Print): 1386-2073
ISSN (Online): 1875-5402

(Research Papers) Time-Resolved Fluorescence Measurements Using Microlens Array and Area Imaging Devices

Author(s): Susanne Merk, Achim Lietz, Margareta Kroner, Martin Valler and Ralf Heilker

Volume 7, Issue 1, 2004

Page: [45 - 54] Pages: 10

DOI: 10.2174/138620704772884814

Price: $65

Abstract

Time-resolved fluorescence (TRF) assay formats are frequently used technologies in highthroughput screening. In this article, we have characterised the novel Plate::Vision2 96-microlens array reader (Carl Zeiss Jena GmbH, Germany) and compared it to the novel LEADseeker Generation IV multimodality imaging system (LEADseeker Gen IV;; Amersham Biosciences UK Ltd., UK) for applications in the TRF mode. In europium measurements using the TRF mode, the Plate::Vision displayed a limit of detection for europium of approximately 3 pM, which was comparable to two established TRF readers, the Discovery and the Victor V (both PerkinElmer Life Sciences Inc., USA). The LEADseekers limit of detection only extended down to europium concentrations of approximately 10 pM in these experiments. For TRF resonance energy transfer (TR-FRET) experiments, a europium-biotin (Eu-biotin) conjugate was titrated with a streptavidin-allophycocyanin (SA-APC) conjugate. The Plate::Vision produced Z values larger than 0.5 for the acceptor fluorophor emission with concentrations of Eu-biotin as low as 3 nM combined with 175 pM SA-APC. To achieve Z values of at least 0.5 with the LEADseeker, concentrations of 10 nM Eu-biotin combined with SA-APC of at least 0.8 nM were required. In a drug screening application using TR-FRET, the energy transfer from a europium-labelled protein X (Eu-protein X) to a complex of biotinylated peptide Y with SA-APC was measured. Using the Plate::Vision, a Z factor larger than 0.5 for the acceptor fluorophor emission was only obtained for a Eu-protein X concentration of at least 10 nM in combination with biotinylated peptide Y / /SA-APC at saturating concentrations. Both the Plate::Vision and the LEADseeker show good quality results for applications in the TRF mode and enable an increased throughput based on their shortened measurement time in comparison to classic photomultiplier tube-based readers.

Keywords: time-resolved fluorescence, microlens array, imaging


Rights & Permissions Print Cite
© 2024 Bentham Science Publishers | Privacy Policy